Fluke 50 Series II Thermometers

Back

Features

 

The new Fluke 50 Series II contact thermometers offer fast response and laboratory accuracy (0.05% + 0.3?C) in a rugged, handheld test tool.

Fluke 50 Series II products features:

  • Large backlit dual display shows any combination of T1, T2 (52 and 54 only), T1-T2 (52 and 54 only), plus MIN, MAX, or AVG
  • Relative time clock on MIN, MAX, and AVG provides a time reference for major events
  • Electronic Offset function allows compensation of thermocouple errors to maximize overall accuracy
  • Measures J, K, T, and E-types of thermocouples
  • Readout in ?C, ?F, or Kelvin (K)
  • Splash and dust resistant case protected by impact absorbing holster
  • User-friendly front panel is easy to set up and operate
  • Sleep mode increases battery life; typical 1000-hour battery life
  • Battery door allows easy battery replacement without breaking the calibration seal
  • Optional ToolPak accessory allows the thermometer to hang from any metal object (with the rare earth magnet) or secure around a pipe (with hook-and-loop straps) for hands-free operation.
  • 3 year warranty

The Fluke 53 Series II and 54 Series II include all the features listed above plus:

  • Data Logging up to 500 points of data with user adjustable recording interval
  • Additional thermocouple types R, S, and N (for a total of 7 different types)
  • Real time clock captures the exact time of day when events occur
  • Recall function allows logged data to be easily reviewed on the meter display
  • IR communication port allows data to be exported to optional FlukeView Forms? Temperature PC software for further analysis and graphing

 

Specifications:

Measurement accuracy Above -100 °C (-148 °F):
J, K, T, E, and N-type: ±[0.05% + 0.3°C (0.5°F)]*
R and S-type: ±[0.05% + 0.4°C (0.7°F)]*

Below -100 °C (-148 °F):
J, K, E, and N-types: ±[0.20% + 0.3°C (0.5°F)]*
T-type: ±[0.50% + 0.3°C (0.5°F)]
Display resolution 0.1°C / °F / K < 1000°
1°C / °F / K ³ 1000°
Measurement range J-type: -210°C to 1200°C (-346°F to 2192°F)

K-type: -200°C to 1372°C (-328°F to 2501°F)

T-type: -250°C to 400°C (-418°F to 752°F)

E-type: -150°C to 1000°C (-238°F to 1832°F)

N-type: -200°C to 1300°C (-328°F to 2372°F)*

R and S-type: 0°C to 1767°C (32°F to 3212°F)*
Temperature scale ITS-90
Applicable standards NIST-175
Environmental Operating temperature:
-10°C to 50°C (14°F to 122°F)

Storage temperature:
-40°C to 60°C (-40°F to 140°F)

Humidity:
0% to 90%; 0°C to 35°C (32°F to 95°F)
0% to 70%; 0°C to 50°C (32°F to 122°F)
General Unit weight: 400g (14 oz)

Dimensions: 17.3 cm L x 8.6 cm W x 3.8 cm D (6.8 in L x 3.4 in W x 1.5 in D)

Battery: 3 AA batteries; typical 1000-hour life

Certification: CE, CSA, TÜV (pending)

Safety: CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2

* Only the Fluke Models 53 and 54 Series II thermometers are capable of measuring N, R, and S-type thermocouples.